How to Calibrate the Decibel Meter

Place your device next to a handheld sound level meter in steady noise and adjust the calibration offset until both show the same number. Without a reference meter you can only make a rough correction.

Why calibration is needed

A browser cannot see sound pressure directly. It sees digital samples from the microphone, and every phone and laptop converts the same sound into a different signal level. Our meter measures that signal in decibels relative to digital full scale (dBFS) and then adds an offset to turn it into an estimated sound pressure level. Calibration means finding the right offset for your device.

Best method: match a reference sound level meter

Any handheld sound level meter works as a reference. A Class 2 or better meter is ideal, but even a basic one beats guessing.

  1. Make steady noise. Pink noise from a speaker, a fan, an air purifier or a running shower all work. Avoid music and speech, because the level jumps around too much to match. Aim for a moderate level, roughly 60–80 dB on the reference meter.
  2. Put the two microphones side by side. Hold them a few centimeters apart, at the same distance from the source and pointing the same way. Stand at least a meter from the source and away from walls, and keep your hands off both microphones.
  3. Use the same settings on both. Set both to A weighting and Slow response.
  4. Match the readings. Open the calibration control on the meter. Either type in what the reference meter shows, or nudge the offset up or down until the two numbers agree.
  5. Check at a second level. Move closer to or farther from the source. If the two stay within a decibel or two of each other, you are done.

If the readings match at a moderate level but drift apart when it gets louder, your browser is probably still applying automatic gain control. Look for the processing notice on the meter. An offset cannot fix that.

Why A-weighting matters when you compare

Weighting changes the number. A-weighting strongly reduces low frequencies to follow how human hearing responds, C-weighting is nearly flat, and Z applies no weighting at all. For sound with a lot of bass, such as traffic, ventilation or music, a dBC or dBZ reading can be 10 dB or more above the dBA reading of the very same sound.

Most consumer meters and almost all health guidance use dBA. If your reference is set to dBA and our meter is on Z, you will bake that mismatch into your offset, and every later reading will be wrong by a different amount depending on the sound. Phone microphones are also weakest at low frequencies, the range A-weighting plays down, so the two instruments agree best in dBA. More on dB versus dBA in what is a decibel.

No reference meter? Rough methods

These can remove a large error. They cannot give you a precise calibration. Expect to stay uncertain by 5 dB or more.

Where the calibration is stored

The offset is saved in this browser on this device, using the browser's local storage. It is never sent to us. See the privacy policy.

How to reset

Use Reset to default in the calibration panel to return to the default offset. Clearing the site data for decibelmeter.io in your browser settings has the same effect, and also clears your other saved settings.

For what calibration can and cannot fix, see accuracy.

Sources

Frequently asked questions

Do I need to calibrate the decibel meter?

Not for comparing louder and quieter on the same device. Calibrate if you want the number itself to be close to what a real sound level meter would show.

Can I calibrate without a sound level meter?

Only roughly. You can compare with a familiar sound of approximately known level, but the result may still be off by 5 dB or more.

Where is my calibration saved?

In your browser's local storage on that device. It is not sent to us, and it does not carry over to another device or another browser.

Why did my calibration disappear?

Clearing site data, using a private or incognito window, or switching to a different browser removes or hides the saved offset. Calibrate again in the browser you normally use.

Should I calibrate in dBA or dBC?

Use the same weighting and response on both instruments. If your reference meter is set to dBA and Slow, set our meter to A and Slow before matching them.