How Loud Is an MRI Machine?

Measured inside a 3-tesla MRI scanner, noise averaged about 91 dBA across scan types with maximums near 105 dBA. Clinical scanners can reach as high as 130 dB, which is why hearing protection is always provided.

Measure your room against 130 dB an MRI machine

Press “Measure now” and allow the microphone to compare your surroundings with this level.

0
Threshold of hearing
30
Whisper
60
Conversation
85
Hearing risk (8 h)
120
Threshold of pain
140
Impulse limit
194
Max undistorted sound
An MRI machine: 91–130 dB

Measured inside the scanner bore, at the patient's head.

NIOSH recommended daily limit at 130 dBA: 1 s (85 dBA for 8 h, halved every 3 dB). Calculate a full day.

Measured levels

A 2023 study published in OTO Open measured sound inside 3-tesla and 7-tesla scanners across a set of common imaging sequences:

ScannerMean average level (LAeq)Mean maximum (LAmax)Loudest sequence
3 tesla (clinical)91.4 dBA96.5 dBA104.8 dBA
7 tesla (research)105.6 dBA114.0 dBA121.6 dBA

The authors note that earlier reports put 1.5 T and 3 T scanners as high as 130 dB for some sequences. The answer depends on the scanner and the type of scan, but it is consistently in the range where unprotected exposure is not acceptable.

Where the noise comes from

The knocking and buzzing come from the gradient coils, electromagnets that are switched on and off rapidly to locate the signal in space. Each pulse of current inside the strong main magnetic field makes the coil assembly flex slightly, and the scanner body radiates that vibration as sound. Different scan types switch the gradients in different patterns, which is why the sound changes during one appointment. In the study, the 7 T scanner was about 14 dB louder on average than the 3 T scanner.

The limits

According to the study, the U.S. Food and Drug Administration's criteria for MRI systems are a peak sound pressure level no higher than 140 dB and an average (LAeq) no higher than 99 dBA with hearing protection in place. In workplace terms, NIOSH's recommended limit is 85 dBA for 8 hours, falling to 15 minutes at 100 dBA, and an MRI exam commonly runs 20 minutes to an hour.

Hearing protection during a scan

Can you measure it?

Not with a phone. Phones, watches and laptops contain metal and electronics and must never be taken into the scanner room; the magnet is always on. If you are curious about the level in the control room or waiting area, the decibel meter works fine there. For what these levels mean, see 100 dB, 120 dB and safe decibel levels.

Sources

Frequently asked questions

How many decibels is an MRI?

A 3 T scanner averaged about 91 dBA across sequences in one study, with a maximum of about 105 dBA. Some scans on clinical machines have been reported as high as 130 dB.

Why is an MRI so loud?

The gradient coils are switched on and off rapidly inside a strong magnetic field. The forces make the coil structure vibrate, and the scanner radiates that as sound.

Can an MRI damage your hearing?

With properly fitted earplugs or headphones the exposure is kept within FDA criteria. Without protection the levels are high enough to be a risk, so always use what you are given.