Measured levels
A 2023 study published in OTO Open measured sound inside 3-tesla and 7-tesla scanners across a set of common imaging sequences:
| Scanner | Mean average level (LAeq) | Mean maximum (LAmax) | Loudest sequence |
|---|---|---|---|
| 3 tesla (clinical) | 91.4 dBA | 96.5 dBA | 104.8 dBA |
| 7 tesla (research) | 105.6 dBA | 114.0 dBA | 121.6 dBA |
The authors note that earlier reports put 1.5 T and 3 T scanners as high as 130 dB for some sequences. The answer depends on the scanner and the type of scan, but it is consistently in the range where unprotected exposure is not acceptable.
Where the noise comes from
The knocking and buzzing come from the gradient coils, electromagnets that are switched on and off rapidly to locate the signal in space. Each pulse of current inside the strong main magnetic field makes the coil assembly flex slightly, and the scanner body radiates that vibration as sound. Different scan types switch the gradients in different patterns, which is why the sound changes during one appointment. In the study, the 7 T scanner was about 14 dB louder on average than the 3 T scanner.
The limits
According to the study, the U.S. Food and Drug Administration's criteria for MRI systems are a peak sound pressure level no higher than 140 dB and an average (LAeq) no higher than 99 dBA with hearing protection in place. In workplace terms, NIOSH's recommended limit is 85 dBA for 8 hours, falling to 15 minutes at 100 dBA, and an MRI exam commonly runs 20 minutes to an hour.
Hearing protection during a scan
- You will be given earplugs, headphones or both. Use them every time, even for a short scan.
- The study notes that well-fitted earplugs reduce the level by 10 to 30 dB. Fit matters: a foam plug that is only half inserted gives much less.
- Some scanners offer "quiet" sequences, which earlier research found can cut the noise by 20 to 40 dB at 3 T.
- Tell the technologist if the sound is painful or if you already have tinnitus or sensitive hearing.
Can you measure it?
Not with a phone. Phones, watches and laptops contain metal and electronics and must never be taken into the scanner room; the magnet is always on. If you are curious about the level in the control room or waiting area, the decibel meter works fine there. For what these levels mean, see 100 dB, 120 dB and safe decibel levels.
Sources
- Akbar et al., "Acoustic Noise Levels in High-field Magnetic Resonance Imaging Scanners," OTO Open (2023)
- NIOSH: Noise and Occupational Hearing Loss
- NIDCD: Noise-Induced Hearing Loss
Frequently asked questions
How many decibels is an MRI?
A 3 T scanner averaged about 91 dBA across sequences in one study, with a maximum of about 105 dBA. Some scans on clinical machines have been reported as high as 130 dB.
Why is an MRI so loud?
The gradient coils are switched on and off rapidly inside a strong magnetic field. The forces make the coil structure vibrate, and the scanner radiates that as sound.
Can an MRI damage your hearing?
With properly fitted earplugs or headphones the exposure is kept within FDA criteria. Without protection the levels are high enough to be a risk, so always use what you are given.
Updated